A comparison of blood pressure measurements in newborns

Research output: Contribution to journalArticlepeer-review

Abstract

Blood pressure monitoring is an essential component of neonatal intensive care. We compared invasive and noninvasive (Dinamap, Marquette, and Dash) recordings in newborns and also noninvasive values obtained from upper and lower limbs. Infants' blood pressure was recorded every 6 hours for 72 hours using three noninvasive devices and compared with invasive readings taken simultaneously. Twenty-five babies were enrolled in the study, with birth weights of 560 to 4500 g and gestation 24 + 1to40 + 5 weeks. Three hundred thirty-two recordings were obtained. Comparison between invasive and non- invasive readings revealed that all three noninvasive monitors overread mean blood pressure. There was no significant difference between the cuff recordings obtained from the upper or lower limbs. All three noninvasive devices overestimated mean blood pressure values compared with invasive monitoring. Clinicians may be falsely reassured by noninvasive monitoring. Mean blood pressure values obtained from the upper and lower limb are similar.

Original languageEnglish
Pages (from-to)113-116
Number of pages4
JournalAmerican Journal of Perinatology
Volume26
Issue number2
DOIs
Publication statusPublished - Feb 2009

Keywords

  • Blood pressure
  • Invasive
  • Neonate
  • Noninvasive