Amperometric immunosensors for screening of polycyclic aromatic hydrocarbons in water

Research output: Contribution to journalArticlepeer-review

Abstract

An amperometric immunosensor with low limit detection was developed for the screening of polycyclic aromatic hydrocarbons (PAHs) in water. The system was based on detecting the specific substance using an immunological reaction by measuring the chemical responses to specific antibodies. An integrated biochip with a three electrode system was fabricated. Gold was used as the working electrode with platinum was used as the counter electrode. A modified Ag/AgCl reference electrode was employed to enhance the stability of the immunosensors. Indirect competition enzyme-linked immunosorbent assay (ELISA) was carried out within the electrode using alkaline phosphatase (AP) as the labelled-enzyme. The system shows acceptable reproducibility and good stability. The immunosensor exhibited a wide linear response to PAHs. A limit of detection for this sensor was in the range of 1 to 10 ng ml-1 in aqueous sample.

Original languageEnglish
Article number012023
JournalJournal of Physics: Conference Series
Volume307
Issue number1
DOIs
Publication statusPublished - 2011
Event16th Conference in the Biennial Sensors and Their Applications - Cork, Ireland
Duration: 12 Sep 201114 Sep 2011

Fingerprint

Dive into the research topics of 'Amperometric immunosensors for screening of polycyclic aromatic hydrocarbons in water'. Together they form a unique fingerprint.

Cite this