Analysis of the 16 Environmental Protection Agency priority polycyclic aromatic hydrocarbons by high performance liquid chromatography-oxidized diamond film electrodes

  • Pierre Bouvrette
  • , Sabahudin Hrapovic
  • , Keith B. Male
  • , John H.T. Luong

Research output: Contribution to journalArticlepeer-review

Abstract

The capabilities of using boron-doped diamond (BDD) thin films as electrode materials for analysis of the 16 US Environmental Protection Agency (EPA) priority polyaromatic hydrocarbons (PAHs) after a liquid chromatographic separation were evaluated. The BDD electrode was able to detect all 16 PAHs with high sensitivity due to the low background current and wide potential window. The method provided detection limits ranging from 12-40 nM (3-10 ppb) and repeatable results over consecutive analysis. Calibration curves were linear up to at least 10 μM for all PAHs. The work shows the promising use of diamond as an amperometric detector in high performance liquid chromatography (HPLC), especially for PAHs and other hydrophobic aromatic compounds.

Original languageEnglish
Pages (from-to)248-256
Number of pages9
JournalJournal of Chromatography A
Volume1103
Issue number2
DOIs
Publication statusPublished - 27 Jan 2006
Externally publishedYes

Keywords

  • Amperometric detection
  • High performance liquid chromatography
  • Oxidized diamond electrode
  • Polycyclic aromatic hydrocarbons