TY - JOUR
T1 - Compact Microstrip Antennas with Enhanced Bandwidth for the Implanted and External Subsystems of a Wireless Retinal Prosthesi
AU - Bahrami, Siroos
AU - Moloudian, Gholamhosein
AU - Miri-Rostami, Seyyed Reza
AU - Bjorninen, Toni
N1 - Publisher Copyright:
© 1963-2012 IEEE.
PY - 2021/5
Y1 - 2021/5
N2 - In this communication, two pairs of compact microstrip antennas were introduced for a dual-unit retinal prosthesis operating at two frequencies of 1.45 and 2.45 GHz. For the 2.45 GHz frequency band, a triangular microstrip antenna with shorting pin and 7 mm times6.93 mm times0.63 mm dimensions is explored to enhance the intraocular element bandwidth. Also, a modified rectangular patch (26 mm times24 mm times1.43 mm) with parasitic radiators is employed for the extraocular element. In the 1.45 GHz frequency band, another triangular microstrip intraocular element (6.25 mm times6 mm times0.63 mm) is proposed by etching a series of slots, which results in lower resonance frequency. For the extraocular element at 1.45 GHz, a planar inverted-f antenna (PIFA) (28 mm times24 mm times1.43 mm) with parasitic radiators is introduced. The effectiveness of intraocular and extraocular antennas was evaluated through simulation and further examined by fabricating to characterize the performance in a wireless system inside eye phantom. Specific absorption rate (SAR) analysis of the proposed antennas, results of the simulation, and experimental coupling measurements are presented. Compared with others' work in this problem, with this proposed structure, we could enhance the impedance bandwidth up to 36% without compromising the coupling, which is advantageous for high-resolution retinal prosthesis.
AB - In this communication, two pairs of compact microstrip antennas were introduced for a dual-unit retinal prosthesis operating at two frequencies of 1.45 and 2.45 GHz. For the 2.45 GHz frequency band, a triangular microstrip antenna with shorting pin and 7 mm times6.93 mm times0.63 mm dimensions is explored to enhance the intraocular element bandwidth. Also, a modified rectangular patch (26 mm times24 mm times1.43 mm) with parasitic radiators is employed for the extraocular element. In the 1.45 GHz frequency band, another triangular microstrip intraocular element (6.25 mm times6 mm times0.63 mm) is proposed by etching a series of slots, which results in lower resonance frequency. For the extraocular element at 1.45 GHz, a planar inverted-f antenna (PIFA) (28 mm times24 mm times1.43 mm) with parasitic radiators is introduced. The effectiveness of intraocular and extraocular antennas was evaluated through simulation and further examined by fabricating to characterize the performance in a wireless system inside eye phantom. Specific absorption rate (SAR) analysis of the proposed antennas, results of the simulation, and experimental coupling measurements are presented. Compared with others' work in this problem, with this proposed structure, we could enhance the impedance bandwidth up to 36% without compromising the coupling, which is advantageous for high-resolution retinal prosthesis.
KW - Compact antenna
KW - implantable antenna
KW - microstrip antenna
KW - retinal prosthesis
KW - specific absorption rate (SAR)
UR - https://www.scopus.com/pages/publications/85105575157
U2 - 10.1109/TAP.2020.3025245
DO - 10.1109/TAP.2020.3025245
M3 - Article
AN - SCOPUS:85105575157
SN - 0018-926X
VL - 69
SP - 2969
EP - 2974
JO - IEEE Transactions on Antennas and Propagation
JF - IEEE Transactions on Antennas and Propagation
IS - 5
M1 - 9205621
ER -