Facile barometric calibration of photoluminescence-based oxygen sensors on a standard rotary evaporator system with a digital vacuum pump

  • Liang Li
  • , Rafael D.L. Gaspar
  • , Dmitri B. Papkovsky

Research output: Contribution to journalArticlepeer-review

Abstract

A simple setup and methodology for barometric calibration of quenched phosphorescence oxygen sensors are described, which utilize a standard rotary evaporator (Rotavap) system equipped with a digital vacuum pump. They were demonstrated with a panel of different sensing materials based on PtBP dye, including solid-state sensor coatings on different substrates (membrane inserts, stickers or inner wall of the vacuum flask), as well as liquid sensor formulations. The sensors inside the flask were exposed to different conditions created by the Rotavap system (dry or humid air, water, temperatures ranging 2°C-40°C, residual total pressure), and their intensity, lifetime or phase shift signals were measured with handheld sensor readers Optech (Mocon) and Firesting (PyroScience) in a contact-less manner to generate multi-point calibration curves. Several critical points of the method, which require careful consideration, have been revealed, including sensor mounting inside the Rotavap, its temperature control during calibration, choice of the reader. Finally, a detailed characterization of several sensor types was carried out and their key operational parameters (τ0, KSV, heterogeneity, T-coefficients), and calibration equations for the different conditions were determined and compared to each other and the results of conventional O2 calibration. Overall, barometric calibration method using Rotavap provides a useful alternative to conventional O2 calibration methods, being simple, accurate, convenient and versatile.

Original languageEnglish
Article number136286
JournalSensors and Actuators B: Chemical
Volume418
DOIs
Publication statusPublished - 1 Nov 2024

Keywords

  • Barometric oxygen calibration
  • Lifetime-based sensing
  • Quenched phosphorescence oxygen sensing
  • Sensor characterization
  • Temperature and humidity effects

Fingerprint

Dive into the research topics of 'Facile barometric calibration of photoluminescence-based oxygen sensors on a standard rotary evaporator system with a digital vacuum pump'. Together they form a unique fingerprint.

Cite this