First experimental demonstration of VMAT combined with MLC tracking for single and multi fraction lung SBRT on an MR-linac

  • Prescilla Uijtewaal
  • , Pim T.S. Borman
  • , Peter L. Woodhead
  • , Charis Kontaxis
  • , Sara L. Hackett
  • , Joost Verhoeff
  • , Bas W. Raaymakers
  • , Martin F. Fast
  • University Medical Center Utrecht
  • Elekta AB

Research output: Contribution to journal › Article › peer-review

Abstract

Background and purpose VMAT is not currently available on MR-linacs but could maximize plan conformality. To mitigate respiration without compromising delivery efficiency, MRI-guided MLC tumour tracking was recently developed for the 1.5 T Unity MR-linac (Elekta AB, Stockholm, Sweden) in combination with IMRT. Here, we provide a first experimental demonstration of VMAT + MLC tracking for several lung SBRT indications. Materials and methods We created central patient and phantom VMAT plans (8×7.5 Gy, 2 arcs) and we created peripheral phantom plans (3×18 & 1×34 Gy, 4 arcs). A motion phantom mimicked subject-recorded respiratory motion (A‾=11 mm, f‾=0.33 Hz, drift‾=0.3 mm/min). This was monitored using 2D-cine MRI at 4 Hz to continuously realign the beam with the target. VMAT + MLC tracking performance was evaluated using 2D film dosimetry and a novel motion-encoded and time-resolved pseudo-3D dosimetry approach. Results We found an MLC leaf and jaw end-to-end latency of 328.05(±3.78) ms and 317.33(±4.64) ms, which was mitigated by a predictor. The VMAT plans required maximum MLC speeds of 12.1 cm/s and MLC tracking superimposed an additional 1.48 cm/s. A local 2%/1 mm gamma analysis with a static measurement as reference, revealed pass-rates of 28–46% without MLC tracking and 88–100% with MLC tracking for the 2D film analysis. Similarly, the pseudo-3D gamma passing-rates increased from 22–77% to 92–100%. The dose area histograms showed that MLC tracking increased the GTV D98% by 5–20% and the PTV D95% by 7–24%, giving similar target coverage as their respective static reference. Conclusion MRI-guided VMAT + MLC tracking is technically feasible on the MR-linac and results in highly conformal dose distribution.
Original languageEnglish
Pages (from-to)149-157
Number of pages9
JournalRadiotherapy and Oncology
Volume174
DOIs
Publication statusPublished - Sept 2022
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Lung cancer
  • MLC tracking
  • MR-linac
  • Prediction
  • VMAT

Fingerprint

Dive into the research topics of 'First experimental demonstration of VMAT combined with MLC tracking for single and multi fraction lung SBRT on an MR-linac'. Together they form a unique fingerprint.

Cite this