Abstract
Time-domain diffuse correlation spectroscopy (TD-DCS) is a newly emerging optical technique that exploits pulsed, yet coherent light to non-invasively resolve the blood flow in depth. In this work, we have explored TD-DCS at longer wavelengths compared to those previously used in literature (i.e., 750–850 nm). The measurements were performed using a custom-made titanium-sapphire mode-locked laser, operating at 1000 nm, and an InGaAs photomultiplier as a detector. Tissue-mimicking phantoms and in vivo measurements during arterial arm cuff occlusion in n = 4 adult volunteers were performed to demonstrate the proof of concept. We obtained a good signal-to-noise ratio, following the hemodynamics continuously with a relatively fast (1 Hz) sampling rate. In all the experiments, the auto-correlation functions show a decay rate approximately five-fold slower compared to shorter wavelengths. This work demonstrates the feasibility of in vivo TD-DCS in this spectral region and its potentiality for biomedical applications.
| Original language | English |
|---|---|
| Pages (from-to) | 3377-3380 |
| Number of pages | 4 |
| Journal | Optics Letters |
| Volume | 45 |
| Issue number | 13 |
| DOIs | |
| Publication status | Published - 1 Jul 2020 |
| Externally published | Yes |
Fingerprint
Dive into the research topics of 'In vivo time-domain diffuse correlation spectroscopy above the water absorption peak'. Together they form a unique fingerprint.Cite this