TY - JOUR
T1 - Light beam diaphragm collimation errors and their effects on radiation dose for pelvic radiography
AU - Brookfield, Hannah
AU - Manning-Stanley, Anthony
AU - England, Andrew
N1 - Publisher Copyright:
© 2015 American Society of Radiologic Technologists.
PY - 2015
Y1 - 2015
N2 - Purpose To investigate the range of collimation errors in x-ray rooms and to calculate their possible effects on the radiation dose for anteroposterior pelvic examinations. Methods A collimator test tool was suspended at 3 heights (14, 21, and 28 cm) above the table Bucky in 9 x-ray rooms. Heights corresponded to the typical patient thickness (mean, ± 2 SD) of 67 patients undergoing anteroposterior pelvic radiography. The x-ray beam was visually collimated to the inner boundary of the test tool and exposed to radiation. Differences between the visualized field size and the resultant x-ray field size (corrected for magnification) indicated a collimation error. Next, using a pelvic phantom, minimum textbook collimation was set and then changed and verified to simulate a range of possible collimation errors. Phantom examinations used a standard anteroposterior technique with exposure termination using outer automatic exposure control chambers. Dose area product (DAP) was recorded. Results All but 1 of the 9 x-ray machines had a smaller irradiated area than was visually set. Errors ranged from a 16% reduction in irradiated field size to a slight overirradiation by 0.4%. Assuming that these errors could be larger in other institutions, additional errors with a range of -27% to 18% were simulated. Increases in field size by 1 cm (superiorly/inferiorly) increased the DAP by 5%. Laterally, a 1-cm increase caused a 4% rise in DAP. Increases of 1 cm in both planes raised DAP by 4%. Discussion Within a single clinical department, minimal collimation errors were demonstrated. Further evidence from multiple centers would be beneficial; however, such low incidences might reflect strict legislative requirements governing the use of ionizing radiation. Understanding the magnitude of any error is important, but it is also important to ascertain an error's influence on the effective radiation dose for any given examination. Conclusion Overall, collimation errors were minimal and favored underirradiation. Small collimation errors can affect DAP and are more dose significant in the superior/inferior plane.
AB - Purpose To investigate the range of collimation errors in x-ray rooms and to calculate their possible effects on the radiation dose for anteroposterior pelvic examinations. Methods A collimator test tool was suspended at 3 heights (14, 21, and 28 cm) above the table Bucky in 9 x-ray rooms. Heights corresponded to the typical patient thickness (mean, ± 2 SD) of 67 patients undergoing anteroposterior pelvic radiography. The x-ray beam was visually collimated to the inner boundary of the test tool and exposed to radiation. Differences between the visualized field size and the resultant x-ray field size (corrected for magnification) indicated a collimation error. Next, using a pelvic phantom, minimum textbook collimation was set and then changed and verified to simulate a range of possible collimation errors. Phantom examinations used a standard anteroposterior technique with exposure termination using outer automatic exposure control chambers. Dose area product (DAP) was recorded. Results All but 1 of the 9 x-ray machines had a smaller irradiated area than was visually set. Errors ranged from a 16% reduction in irradiated field size to a slight overirradiation by 0.4%. Assuming that these errors could be larger in other institutions, additional errors with a range of -27% to 18% were simulated. Increases in field size by 1 cm (superiorly/inferiorly) increased the DAP by 5%. Laterally, a 1-cm increase caused a 4% rise in DAP. Increases of 1 cm in both planes raised DAP by 4%. Discussion Within a single clinical department, minimal collimation errors were demonstrated. Further evidence from multiple centers would be beneficial; however, such low incidences might reflect strict legislative requirements governing the use of ionizing radiation. Understanding the magnitude of any error is important, but it is also important to ascertain an error's influence on the effective radiation dose for any given examination. Conclusion Overall, collimation errors were minimal and favored underirradiation. Small collimation errors can affect DAP and are more dose significant in the superior/inferior plane.
UR - https://www.scopus.com/pages/publications/84929156966
M3 - Article
C2 - 25835404
AN - SCOPUS:84929156966
SN - 0033-8397
VL - 86
SP - 379
EP - 391
JO - Radiologic Technology
JF - Radiologic Technology
IS - 4
ER -