Low-dose CT of the abdomen: Evaluation of image improvement with use of noise reduction filters - Pilot study

  • Mannudeep K. Kalra
  • , Michael M. Maher
  • , Dushyant V. Sahani
  • , Michael A. Blake
  • , Peter F. Hahn
  • , Gopal B. Avinash
  • , Thomas L. Toth
  • , Elkan Halpern
  • , Sanjay Saini
  • Massachusetts General Hospital
  • General Electric Medical Systems

Research output: Contribution to journalArticlepeer-review

Abstract

A prospective assessment of improvement in image quality at low-radiation-dose computed tomography (CT) of the abdomen by using noise reduction filters was performed. CT images acquired at standard and 50% reduced tube current were processed with six noise reduction filters and evaluated by three radiologists for image noise, sharpness, contrast, and overall image quality in terms of abdominal organ depiction. Quantitative image noise and contrast-to-noise ratio were measured. Baseline low-dose CT images were significantly worse than standard-dose CT images (P < .05). A statistically significant reduction of noise in low-dose images processed with three filters was noted. In conclusion, use of noise reduction filters decreased image noise at low-dose CT.

Original languageEnglish
Pages (from-to)251-256
Number of pages6
JournalRadiology
Volume228
Issue number1
DOIs
Publication statusPublished - 1 Jul 2003
Externally publishedYes

Keywords

  • Abdomen, CT
  • Computed tomography (CT), image processing
  • Computed tomography (CT), image quality
  • Computed tomography (CT), radiation exposure
  • Radiations, exposure to patients and personnel