Measurement of breast lesion display luminance and overall image display luminance relative to optimum luminance for contrast perception

  • Mohammad Rawashdeh
  • , Warwick Lee
  • , Patrick Brennan
  • , Warren Reed
  • , Mark McEntee
  • , Roger Bourne
  • University of Sydney

Research output: Chapter in Book/Report/Conference proceedingsConference proceedingpeer-review

Abstract

Introduction: To minimize fatigue due to eye adaptation and maximize contrast perception, it has been suggested that lesion luminance be matched to overall image luminance to perceive the greatest number of grey level differences. This work examines whether lesion display luminance matches the overall image and breast tissue display luminance and whether these factors are positioned within the optimum luminance for maximal contrast sensitivity. Methods: A set of 42 mammograms, collected from 21 patients and containing 15 malignant and 6 benign lesions, was used to assess overall image luminance. Each image displayed on the monitor was divided into 16 equal regions. The luminance at the midpoint of each region was measured using a calibrated photometer and the overall image luminance was calculated. Average breast tissue display luminance was calculated from the subset of regions containing of only breast tissue. Lesion display luminance was compared with both overall image display luminance and average breast tissue display luminance. Results: Statistically significant differences (p<0.0001) were noted between overall image display luminance (4.3±0.7 cd/m2) and lesion display luminance (15.0±6.8 cd/m2); and between average breast tissue display luminance (6.8±1.3 cd/m2) and lesion display luminance (p<0.002). Conclusions: Lesion luminance was significantly higher than the overall image and breast tissue luminance. Luminance of lesions and general breast tissue fell below the optimum luminance range for contrast perception. Breast lesion detection sensitivity and specificity may be enhanced by use of brighter monitor displays.

Original languageEnglish
Title of host publicationMedical Imaging 2011
Subtitle of host publicationImage Perception, Observer Performance, and Technology Assessment
DOIs
Publication statusPublished - 2011
Externally publishedYes
EventMedical Imaging 2011: Image Perception, Observer Performance, and Technology Assessment - Lake Buena Vista, FL, United States
Duration: 16 Feb 201117 Feb 2011

Publication series

NameProgress in Biomedical Optics and Imaging - Proceedings of SPIE
Volume7966
ISSN (Print)1605-7422

Conference

ConferenceMedical Imaging 2011: Image Perception, Observer Performance, and Technology Assessment
Country/TerritoryUnited States
CityLake Buena Vista, FL
Period16/02/1117/02/11

Keywords

  • Breast lesion
  • Contrast perception
  • Display luminance
  • Eye adaptation
  • Mammogram

Fingerprint