RF performance of a 418-MHz radio telemeter packaged for human vaginal placement

Research output: Contribution to journalArticlepeer-review

Abstract

The electrical and communication performance of a 0.8-μW UHF temperature telemeter designed for human vaginal placement is discussed; a solenoidal loop antenna was used, occupying a volume of 0.1 cm3. In situ, measured power absorption was between 19-25 dB, resulting in an effective operating range of 10 m. Capacitive loading lowered the antenna's resonant frequency by 1.4% and there was a significant polarization change in the radiated output.

Original languageEnglish
Pages (from-to)427-430
Number of pages4
JournalIEEE Transactions on Biomedical Engineering
Volume44
Issue number5
DOIs
Publication statusPublished - 1997
Externally publishedYes

Keywords

  • Biomedical telemetry
  • SPICE
  • UHF antennas

Fingerprint

Dive into the research topics of 'RF performance of a 418-MHz radio telemeter packaged for human vaginal placement'. Together they form a unique fingerprint.