The reliability of point-of-care prothrombin time testing. A comparison of CoaguChek S and XS INR measurements with hospital laboratory monitoring

Research output: Contribution to journalArticlepeer-review

Abstract

The development of point-of-care (POC)testing devicesenablespatients to test their own international normalized ratio (INR) at home. However, previous studies have shown that when compared with clinical laboratory values, statistically significant differences may occur between the two methods of INR measurement. The aim of this study was to evaluate the accuracy of the CoaguChek S and XS POC meters relative to clinical laboratory measurements. As part of a randomized, crossover patient self-testing (PST) study at Cork University Hospital, patients were randomized to 6 months PST or 6 months routine care by the antico- agulation management service. During the PST arm of the study, patients measured their INR at home using the CoaguChek S or XS POC meter. External quality control was performed at enrolment, 2 months and 4 months by comparing the POC measured INR with the laboratory determined value. One hundred and fifty-one patients provided 673 paired samples. Good correlation was shown between the two methods of determination (r = 0.91), however, statistically significant differences did occur. A Bland-Altman plot illustrated good agreement of INR values between 2.0 and 3.5 INR units but there was increasing disagreement as theINR rose above 3.5.Eighty-seven per cent of all dual measurements were within the recommended 0.5 INR units of each other. This study adds to the growing evidence that POC testing is a reliable and safe alternative to hospital laboratory monitoring but highlights the importance of external quality control when these devices are used for monitoring oral anticoagulation.

Original languageEnglish
Pages (from-to)e26-e33
JournalInternational Journal of Laboratory Hematology
Volume32
Issue number1 PART.1
DOIs
Publication statusPublished - Feb 2010

Keywords

  • Anticoagulation
  • International normalized ratio
  • Patient self-testing
  • Point-of-care
  • Quality control

Fingerprint

Dive into the research topics of 'The reliability of point-of-care prothrombin time testing. A comparison of CoaguChek S and XS INR measurements with hospital laboratory monitoring'. Together they form a unique fingerprint.

Cite this